The Impact of Differential Atrioventricular Delays on Arterial Distensibility Measured by Pulse Wave Velocity in Patients with Dual-chamber Pacing
PDF
Cite
Share
Request
Original Research
P: 88-93
March 2023

The Impact of Differential Atrioventricular Delays on Arterial Distensibility Measured by Pulse Wave Velocity in Patients with Dual-chamber Pacing

Bagcilar Med Bull 2023;8(1):88-93
1. University of Health Sciences Turkey, Prof. Dr. Cemil Taşcıoğlu City Hospital, Clinic of Cardiology, İstanbul, Turkey
2. İstanbul University-Cerrahpaşa, Institute of Cardiology, İstanbul, Turkey
No information available.
No information available
Received Date: 02.03.2023
Accepted Date: 04.03.2023
Publish Date: 10.03.2023
PDF
Cite
Share
Request

ABSTRACT

Conclusion:

Prolongation of AV delay interval increased carotid-femoral PWV values thereby decreased arterial distensibility in patients with DDD pacing. Therefore, optimum AV delay adjustment may provide better hemodynamics in these patients.

Results:

There was a statistically obvious positive correlation between carotid-femoral PWV and prolongation of AV delay interval (p<0.001). According to gender analysis; weight, height and waist-hip ratio values were significantly higher in male group. Additionally, there was not a statistically significant difference between groups in terms of age, body mass index, mean, systolic and diastolic blood pressure, pulse pressure and PWV measured at differential AV delay intervals.

Method:

A total of 40 patients with DDD pacing were enrolled in our prospective study. PWV was measured for each AV delay (100 ms, 150 ms and 200 ms respectively) with 10 minute resting intervals. Pacing was programmed at least 10 beats/minute above the resting heart rate. PWV was automatically calculated by using the device, Complior Colson (France), which allows online pulse wave recording according to this formula: PWV(m/s) = distance (meter)/transit time (second).

Objective:

Atrioventricular (AV) delay optimization improves hemodynamics and clinical parameters of patients treated with dual-chamber (DDD) pacemakers. Pulse wave velocity (PWV) is an index of arterial stiffness and a marker of cardiovascular events. Increased levels of PWV is an indicator of diffuse atherosclerosis and rigid arteries. We aimed to investigate the impact of differential AV delay intervals (100, 150 and 200 ms respectively) on arterial distensibility measured by carotid-femoral PWV in patients with DDD pacing.

Introduction

Newer generation dual-chamber (DDD) pacemakers have been widely used as they are considered to be more physiological (1). However, DDD pacemakers require more timers than other devices as they include several types of intervals, such as the basic and atrioventricular (AV) delay intervals.

Arterial stiffness is an indicator of atherosclerosis and caused by loss of elasticity and thickening of the arterial wall. Increased arterial stiffness is not only an indicator of vascular aging, but also a predictor of target organ damage and increased cardiovascular events (2). Due to the insidious nature of the atherosclerotic process, early recognition of arterial changes and functional and/or structural lesions may help to identify patients with high risk for clinical complications.

Pulse wave velocity (PWV) is one of the most important parameters used to evaluate elastic properties of large arteries (3,4). Also, using PWV measurements provides an easy and reproducible method to assess arterial stiffness (5). PWV is measured by using two ultrasound or pressure sensitive transducers fixed transcutaneously on the trace of a couple of artery at a certain distance from each other (such as carotid-femoral or brachial-radial arteries) (3). Increase in PWV values, which is defined as the velocity of the pressure wave over an arterial segment, is an important indicator of the diffuse atherosclerotic process (6). Arterial PWV is inversely proportional to arterial compliance and PWV propagation time (PWVPT) (3,4).

In addition, increased heart rate causes an increase in arterial PWV (7); thereby a decrease in arterial distensibility and compliance. Previous studies have shown that changes in AV delay interval affect both the left ventricular filling time and stroke volume of the patients (8,9). It is unknown whether other different factors such as AV conduction time affects PWV or not in case of a fixed heart rate.

Therefore, we assumed that PWV values calculated at different AV delay intervals may guide optimal programming to provide better hemodynamics in patients with DDD pacemakers. Due to the lack of data regarding this issue, we aimed to investigate the effect of differential AV delay intervals (100, 150 and 200 ms, respectively) on arterial distensibility measured by the carotid-femoral PWV in patients with DDD pacing.

Materials and Methods

Fourty consecutive patients with DDD pacemakers who were admitted to our outpatient clinic between January to July 2015, were recruited in our single-center and prospective clinical study. Written informed consent was taken from each participant. Our study was conducted in concordance with the Declaration of Helsinki and approved by the Local Ethics Committee of İstanbul University (09/07/2014, no: B.08.06.YOK.2.I.U.E.50.0.05.00/10).

None of the patients had a pacemaker other than DDD mode. Exclusion criteria included patients with coronary artery disease, cerebrovascular disease, peripheral artery disease, heart failure, liver or kidney failure, atrial fibrillation and patients using non-steroidal anti-inflammatory or steroid drugs during the study. We also excluded patients with body mass index (BMI) ≥35 kg/m² and waist-hip ratio (WHR) ≥1.

Patients were asked to rest in a supine position for 10 minutes before starting the measurements. Following each 10 minute period of haemodynamic stabilization, AV delay intervals were gradually prolonged and programmed as 100, 150 and 200 ms, respectively. Three different AV delays were used for testing by AV sequential pacing. Carotid-femoral artery PWV and blood pressure (BP) were measured by same researcher. Pacing was programmed at a minimum of 10 beats/minute above the resting heart rate. Also, pacing rate was further increased when continuous capture could not be obtained (8). Weight measurements of the patients were made by scale, whereas height measurements by meter. BMI was calculated by dividing the weight in kilogramme by the height calculated in square meters (kg/m2). The waist circumference (cm) was measured from the mid point of the distance between the lowest rib and the crista iliaca when the patient was in upright position. The hip circumference (cm) was measured from the major trochanter femoris level. WHR was calculated by dividing waist circumference by hip circumference. Arterial BP was measured with a standard mercury manometer in supine position following a 10-minute period of rest. Systolic BP (SBP) was accepted as the value at which Korotkoff sounds were heard first and diastolic BP (DBP) as the value at which sounds disappeared (phase 5). Pulse pressure (PP) was calculated by subtracting DBP from SBP. Mean BP (MBP) was determined using the formula MBP = SBP + 2xDKB/3.PWV measurements were taken immediately after obtaining the brachial BP. Aortic PWV was calculated using the Complior device (Createch Industrie, France), which allows automatic online pulse wave recording and automatic calculation of PWV (3). Common carotid and femoral artery pressure waveforms were measured non-invasively using a pressure sensitive transducer TY-306 Fukuda (Fukuda, Tokyo, Japan). An average of ten repeated measurements was taken for final analysis. PWV was automatically calculated with the formula ΔD/Δt [ΔD: distance traveled by the pulse wave on the body surface (meters) between two recording points, Δt: Pulse wave transit time (seconds) automatically determined by the Complior device].

Statistical Analysis

Statistical analysis was performed by using the SPSS version 15.0 (SPSS Inc., Chicago, Illinois, USA) packet program. Continuous variables were reported as mean ± standard deviation while categorical variables were presented as numbers and percentages. The significance between different measurements of the same participant was determined by using the non-parametric Friedman test. If significance was detected in the Friedman test, the non-parametric Wilcoxon test was used to determine the source of significance. As AV delay parameters were non-normally distributed, Friedman and Wilcoxon tests were used. The non-parametric Mann-Whitney U test was used for comparison between the female and male groups. Student t-test was used for normally distributed data. The relationship between PWV and other variables was evaluated by Pearson correlation test. Values of p<0.05 were considered statistically significant.

Results

Of the 40 cases included in the study, 21 (52.5%) were male and 19 (47.5%) were female. Demographic characteristics and anthropometric measurements of the patients were shown in Table 1. Type II diabetes mellitus was present in 7 (36.8%) of 19 female cases and 5 (23.8%) of 21 male cases. Hypertension was found in 15 (78.9%) patients among women and 11 (52.3%) patients among men. Eight (42.1%) female and five (23.8%) male patients had hyperlipidemia.

Table 1

The change in BP with respect to increasing AV delay intervals was shown in Table 2. No statistically significant difference was found between SBP, DBP, MBP and PP values measured at differential AV delay intervals (p=0.105, p=0.264, p=0.449 and p=1.000, respectively).

Table 2

A statistically significant increase in PWV and a decrease in PWVPT were detected with the prolongation of AV delay intervals (p<0.001 and p<0.001, respectively) and graphically shown in Figure 1A and 1B. When the groups were compared among themselves by the Wilcoxon test; a statistically significant difference was found between PWV 100-150 ms, 100-200 ms and 150-200 ms (for PWV 100 ms vs. 150 ms, p<0.001; PWV 100 ms vs. 200 ms, p<0.001; PWV 150 ms vs. 200 ms p=0.001). Similarly, statistically significant results were obtained in the comparison between the groups in terms of PWVPT (for PWVPT 100 ms vs. 150 ms, p=0.002; PWVPT 100 ms vs 200 ms, p<0.001; PWVPT 150 ms vs. 200 ms p<0.001, respectively). A strong positive correlation (p<0.001, r=0.590) was found between age and the PWV values taken at 100 ms. Likewise, there was a strong positive correlation between age and PWV values taken at 150 ms and 200 ms (respectively p<0.001, p<0.001; r=0.611, r=0.562).

According to gender analysis that was given in Table 3; height, weight and WHR values of men were significantly higher than those of women (p<0.001, p=0.016 and p=0.001, respectively). Also, there was no statistically significant difference between women and men in terms of age, BMI, SBP, DBP, MBP, PP, PWV and PWVPT values measured at differential AV delay intervals (100, 150 and 200 ms, respectively).

Table 3

Discussion

To the best of our knowledge, this is the first study to investigate the effect of different AV delay intervals on arterial distensibility measured by carotid-femoral PWV in patients with DDD pacing. Our study demonstrated a statistically significant increase in arterial PWV and thereby a decrease in PWVPT with gradually increasing AV delay intervals (100, 150 and 200 ms). In this respect, the results of our study suggest that prolongation of AV conduction time may increase arterial stiffness in patients with classical DDD pacemakers.

Nishimura et al. (10) examined the patients with dilated cardiomyopathy into two groups according to PR intervals whether above or below 200 milliseconds. A significant increase in cardiac output of patients in long PR interval group was noted with AV interval optimization. Likewise, Manisty et al. (8) evaluated hemodynamic changes occurred at different AV delay intervals (40-120 ms) in 19 patients with permanent pacemaker, and observed an acute improvement in hemodynamic parameters (SBP, DBP, MBP, PP and stroke volume) as the AV delay interval approached from non-physiological (40 ms) to physiological values (120 ms). Contrary to this study, no significant difference was found in our study in terms of hemodynamic parameters. This may be due to the fact that we recorded our measurements in a more physiological range of AV delay, such as 100 ms to 200 ms.

Patients whose arterial PWV values increase may have worse cardiac functions due to the increased workload in front of the left ventricle. Also, decrease in arterial distensibility may lead to decreased effective coronary blood flow and increased myocardial ischemia (11). Various studies have suggested that increased PWV values may be related to increased adverse cardiovascular events and mortality (3,6,11,12).

PWV may be affected by different parameters ​​such as age, heart rate, SBP, DBP, MBP, and PP (12,13). In the present study, a significant relationship was found between PWV and age. As the age increases, the elastic tissue of the aorta gradually decreases and therefore stiffness increases. Additionally, increasing the right ventricular pacing rate may increase arterial stiffness. Yıldız et al. (14) compared 17 patients with classical DDD pacemakers with healthy subjects in terms of arterial distensibility determined by PWV, and found an increase in arterial PWV and a decrease in distensibility in patients with pacemakers in terms of gender and age compared to the control group. Similarly, Wilkinson et al. (15) reported a linear relationship between heart rate and arterial stiffness in patients with permanent atrial or dual pacemakers. Krishnamoorthy et al. (16) divided 101 patients with DDD pacemakers into two groups according to the presence of atrial high-rate episodes. An increase in arterial stiffness was detected in the group with atrial high velocity episodes. These studies were not conducted at a constant heart rate unlike our study, in which the effect of heart rate was excluded.

The effect of differential AV delay intervals on arterial stiffness still remains controversial. Differential AV delay intervals may lead to hemodynamic adaptation and lengthen the time of systole or diastole (17). The prolongation of systolic phase, occured by AV conduction programming, may cause an increase in left ventricular stroke volume and thus in the amount of blood passed and consequently an increase in aortic pressure. Chan et al. (18) investigated the changes in vascular function with PR (AV delay interval) prolongation in 88 healthy individuals. As the AV interval prolonged, a decrease in flow-mediated dilatation and an increase in PWV was observed. This also indicates that PR prolongation may a negative effect on arterial mechanics.

Evaluation of arterial stiffness is increasingly important in patients with high cardiac risk such as the presence of a permanent pacemaker. Based on this, measuring arterial expandability besides hemodynamic changes when programming the pacemaker seems to be reasonable in these patients. If their devices are programmed for the optimum AV delay interval, these patients may experience improvement in large artery mechanics such as arterial expansion.

Study Limitations

There were some limitations to be noted in our study. First of all, our results were based on a single-center study and relatively small number of patients were analyzed. Secondly, most of the patients have coexistent medical conditions such as diabetes mellitus and hypertension; therefore our results might not accurately reflect endothelial dysfunction associated with AV delay interval prolongation. Antihypertensive and antidiabetic drugs also may effect our results.

Conclusion

In conclusion, the results of this study indicated that lengthening AV delay interval increased carotid-femoral PWV in patients with DDD pacemakers. Thus, optimum AV delay adjustment may provide better hemodynamics in these patients. Nevertheless, further prospective studies are needed to confirm our findings in a larger cohort of patients.

Acknowledgments

The authors thank to Urfan Jafarov for his special attribution for the study.

Ethics

Ethics Committee Approval: Fourty consecutive patients with DDD pacemakers who were admitted to our outpatient clinic between January to July 2015, were recruited in our single-center and prospective clinical study. Our study was conducted in concordance with the Declaration of Helsinki and approved by the Local Ethics Committee of İstanbul University (09/07/2014, no: B.08.06.YOK.2.I.U.E.50.0.05.00/10).

Informed Consent: Written informed consent was taken from each participant.

Peer-review: Externally and internally peer-reviewed.

Authorship Contributions

Concept: B.B.K., G.Ç., M.Y., Design: B.B.K., G.Ç., M.Y., Data Collection or Processing: B.B.K., G.Ç., Analysis or Interpretation: B.B.K., G.Ç., M.Y., Literature Search: B.B.K., G.Ç., Writing: B.B.K., G.Ç.

Conflict of Interest: No conflict of interest was declared by the authors.

Financial Disclosure: The authors declared that this study received no financial support.

References

1
Ochagavía Calvo A, Baigorri González F. [Selection of pacemaker stimulation mode]. Med Intensiva 2006;30(5):218-222.
2
Sakuragi S, Abhayaratna WP. Arterial stiffness: methods of measurement, physiologic determinants and prediction of cardiovascular outcomes. Int J Cardiol 2010;138(2):112-118. 
3
Pichler G, Martinez F, Vicente A, Solaz E, Calaforra O, Redon J. Carotid-femoral pulse wave velocity assessment by two different methods: implications for risk assessment. J Hypertens 2015;33(9):1868-1875.
4
Imura R, Yamamoto K, Kanamori K, Mikami T, Yasuda H. Non invasive ultrasonic measurement of the elastic properties of the human abdominal aorta. Cardiovasc Res 1986;20(3):208-214.
5
Asmar R. Effects of pharmacological intervention on arterial stiffness using pulse wave velocity measurement. J Am Soc Hypertens. 2007;1(2):104-112.
6
Laurent S, Boutouyrie P, Asmar R, Gautier I, Laloux B, Guize L, et al. Aortic stiffnessis an independent predictor of all-cause and cardiovascular mortality in hypertensive patients. Hypertension 2001;37(5):1236-1241.
7
Albaladejo P, Copie X, Boutouyrie P, Laloux B, Déclère AD, Smulyan H, et al. Heart rate, arterial stiffness, and wave reflections in paced patients. Hypertension 2001;38(4):949-952.
8
Manisty CH, Al-Hussaini A, Unsworth B, Baruah R, Pabari PA, Mayet J, et al. The acute effects of changes to AV delay on BP and stroke volume: potential implications for design of pacemaker optimization protocols. Circ Arrhythm Electrophysiol 2012;5(1):122-130.
9
Shinbane JS, Chu E, DeMarco T, Sobol Y, Fitzpatrick AP, Lau DM, et al. Evaluation of acute dual-chamber pacing with a range of atrioventricular delays on cardiac performance in refractory heart failure. J Am Coll Cardiol 1997;30(5):1295-1300.
10
Nishimura RA, Hayes DL, Holmes DR, Tajik AJ. Mechanism of hemodynamic improvement by a dual chamber pacing for severe left ventricular dysfunction. J Am Coll Cardiol 1995;25(2):281-288.
11
Graham RM, Perez DM, Hwa J, Piascik MT. Alpha-1 Adrenergic Receptor Subtypes. Molecular Structure, Function and Signaling. Circ Res 1996;78(5):737-749.
12
Wang F, Ye P, Luo L, Xiao W, Qi L, Bian S, et al. Association of serum lipids with arterial stiffness in a population-based study in Beijing. Eur J Clin Invest. 2011;41(9):929-936.
13
Asmar R. Factors influencing pulse wave velocity. Arterial stiffness and pulse wave velocity. Elsevier, 1999:57-88.
14
Yıldız M, Yıldız BŞ, Şeker M, Hasdemir H. The Effects of Traditional Dual Chamber Cardiac Permanent Pacemaker on Arterial Distensibility Using CarotidFemoral (Aortic) Pulse Wave Velocity in Patients with Angiographically Normal Coronary Arteries. Kosuyolu Kalp Derg 2012;15(3):93-99.
15
Wilkinson IB, MacCallum H, Flint L, Cockcroft JR, Newby DE, Webb DJ. The influence of heart rate on augmentation index and central arterial pressure in humans. J Physiol 2000;525(Pt 1):263-270.
16
Krishnamoorthy S, Khoo CW, Lim HS, Lip GY. Predictive value of atrial high-rate episodes for arterial stiffness and endothelial dysfunction in dual-chamber pacemaker patients. Eur J Clin Invest 2014;44(1):13-21.
17
Krychtiuk KA, Nürnberg M, Volker R, Pachinger L, Jarai R, Freynhofer MK, et al. Effects of AV-delay optimization on hemodynamic parameters in patients with VDD pacemakers. Wien Klin Wochenschr 2014;126(9-10):270-277.
18
Chan YH, Siu CW, Yiu KH, Li SW, Lau KK, Lam TH, et al. Abnormal vascular function in PR-interval prolongation. Clin Cardiol 2011;34(10):628-632.